6
•
Operation
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Page 95
6.6
Important Operating Parameters
The commands and parameters described in the table should be
considered for simple routine operation of the device. You can usually
access these parameters from the Chromeleon user interface.
If a parameter listed below is not available in the Chromeleon software,
consider updating the firmware and Chromeleon version.
TIP
The Instrument Method Wizard provides different parameter view
modes. Depending on the parameters to be set, you can select the
desired view mode (for example,
Easy
or
Advanced
).
For more information, refer to
Chromeleon Help and documents
.
General Parameters
Parameters
Description
UV lamp
VIS lamp (optional)
Turn on the UV and/or VIS lamp before starting an
analysis (
UV_Lamp = On, Visible_Lamp =On
). This setting
allows turning on and off the UV and VIS lamp of the
device.
To turn on the UV lamp automatically after power on, set
AutoactivateUV_Lamp
to
On
.
To turn on the VIS lamp automatically after power on, set
AutoactivateVIS_Lamp
to
On
.
UV_VIS
The
UV_VIS
signal channels (
UV_VIS_x
) that are available
for data acquisition are listed in the dialog box for the
detector in the Instrument Configuration Manager. Select
the required signal channels.
The device can record several signal channels
simultaneously, thus at several different wavelengths (for
the number of available channels, see
Acquisition On
Acquisition Off
Start or stop the data acquisition.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...