7
•
Maintenance and Service
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Page 133
6. Push the isopropanol in the syringe into the flow cell. Make sure that
no residual air in the syringe is pushed into the flow cell.
7. Disconnect the waste line from the flow cell outlet.
8. Disconnect the flushing adapter from the flow cell inlet.
9. Install the flow cell plugs to the flow cell inlet and flow cell outlet.
10. Unscrew the flushing adapter from the syringe.
11. Make sure that the adapter and the syringe are empty and dry.
12. Put the protection cap onto the Viper adapter.
13. Pack and store the components of the kit in their original packaging.
14.
If the flow cell is to be stored in its packaging
Remove the flow cell from the device and store it (see
).
7.7.2.1
Performing a Manual Injection
A sample can also be injected manually using the flushing and injection
kit.
Parts required
Flushing and injection kit for flow cells, including a syringe and a Viper
adapter
NOTICE
Particulate matter from the eluate can deposit in the flow cell and clog
it.
•
Be careful not to inject any particulate matter into the flow cell.
•
Only experienced users should perform a manual injection.
The manual injection requires setting suitable detection parameters,
starting acquisition, performing an autozero and data interpretation
afterward. The related injection procedures of autozero injection,
sample injection and flushing are similar to the flushing procedure
described in
Preparing the Flow Cell for Storage (
.
Observe the following guidelines when performing a manual injection:
•
Avoid injecting any particulate matter into the flow cell.
•
All solvents used must be miscible.
•
Flush the sample completely out of the flow cell after measurement.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...