8
•
Troubleshooting
Page 164
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Message and Code
Description and Remedial Action
Code 3156 xx failed. D-alpha line not
found - check flow cell.
xx = Wavelength calibration / wavelength validation
Identification of D-alpha line during calibration/validation failed.
•
Repeat wavelength calibration (see
•
).
•
If the message still appears, check whether the UV lamp is
installed properly and whether it is turned on.
•
Replace the UV lamp (see
•
If the wavelength calibration and the wavelength validation still
fail, contact Technical Support.
Code 3160 Internal process timed out.
•
Repeat the previous operation. When the message appears
again, turn off the module. Wait for 5 seconds and turn on the
module again.
•
Upgrade the firmware, if available.
Code 3162
xx over-temperature - emergency
standby entered now.
xx = Affected sensor (system or lamp house)
•
Turn off the module and let it cool down.
•
Check for cooling air blockage.
•
Lower the environmental temperature.
Code 3198 Invalid spectral data. Check
flow cell.
Invalid spectral data for absorption calculation.
•
Make sure that the mobile phase does not show excessive
absorption for the selected channel wavelength(s).
•
).
•
If the message still appears, check whether the lamps, especially
the UV lamp, are installed properly.
•
Code 3199 Invalid auto-zero spectral
data. Check flow cell.
Invalid spectral data for absorption calculation.
•
Make sure that the mobile phase does not show excessive
absorption for the selected channel wavelength(s) during
autozero.
•
Select a start time for autozero when the flow cell is only filled
with eluent but not with eluate.
•
).
•
If the message still appears, check whether the lamps, especially
the UV lamp, are installed properly.
•
Code 3201 No or unknown xx installed.
xx = Flow cell, UV/VIS (optional) lamp or diagnostic cell
No flow cell, UV/VIS (optional) lamp or diagnostic cell installed or
without ID chip. Install the mentioned component (flow cell,
diagnostic cell or lamp) with ID chip.
Code 3202 Acquisition aborted (flow cell
removed).
Reinstall the flow cell or diagnostic cell. Restart data acquisition.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...