11.
If you install a VIS lamp for the first time and the detector is operated
with Chromeleon, activate the lamp in Chromeleon:
Only then, the VIS lamp can be ignited from Chromeleon.
a) Right-click the detector in the Instrument Configuration
Manager.
b) Select
Properties
on the menu.
c) On the
Detector
page, enable the
VIS Lamp
check box.
12. After installing a new lamp, increased noise and strong baseline
fluctuations may occur. Before beginning an analysis or performing a
wavelength validation, allow the new lamp to run until the noise is
reduced and the baseline is stable which typically takes 24 hours.
13. Perform a wavelength validation (see
).
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...