7
•
Maintenance and Service
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Page 139
7.8.2
Removing the Diagnostic Cell
Preparations
1. Loosen the two screws of the flow cell cover, remove the cover and
keep the cover to close the flow cell opening after this procedure.
Follow these steps
1. Carefully pull out the diagnostic cell from the opening in the device.
CAUTION—UV Radiation
The deuterium lamp emits UV radiation that is harmful to eyes and skin.
Do not look and do not reach into the flow cell opening while no flow
cell or diagnostic cell is installed and the UV lamp is turned on.
2. Store the diagnostic cell in its packaging.
3.
If a flow cell is to be installed afterward:
Install the flow cell to the flow cell opening on the device (see
4. Mount the flow cell cover so that the slot in the flow cell cover is
located on the bottom right (see image). Make sure that the
capillaries run through the slot in the flow cell cover.
Figure 32: Fixing the flow cell cover (here: flow cell with connection unit)
5. Fasten the two screws of the flow cell cover hand-tight.
NOTICE
The flow cell opening on the device is sensitive to dust and debris.
If no flow cell is installed to the device, close the flow cell opening with
the cover for the flow cell opening.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...