5
•
Installation
Page 74
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Resolving Leakage of Viper Fittings with Knurls
1. Tighten the connection a little more.
2. If leakage continues, remove the capillary.
3. Clean the capillary ends carefully by using a lint-free tissue wetted
with isopropanol.
4. Reinstall the capillary.
5. If the connection continues to leak, install a new Viper capillary.
5.8.5
Flow Connections to the Flow Cell
In addition, observe the guidelines for proper connection of the flow cell
in
General Information and Guidelines (
.
Parts required
•
Inlet capillary
•
Detector waste line
For instructions on connecting the waste line, follow the steps in
Connecting the Detector Waste Line (
.
Tools required
For the detector waste line:
Tubing cutter (optional)
Preparations
1. Flush the system modules and capillaries upstream of the device to
waste before you connect the flow cell to the system flow path.
Refer to the
Vanquish System Operating Manual
.
2. On the flow cell, remove the plugs from the flow cell inlet and
outlet.
TIP
Store the plugs of the flow cell, for example in the flow cell
packaging, to have them easily available when storing or shipping the
flow cell.
3. Install the partition panel plugs (if not present yet). See
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...