6.7.2.5
Baseline Behavior
If wavelengths are switched during a data acquisition, the
chromatogram may show baseline jumps.
You can define how the baseline behaves in these cases:
•
Zero
: sets the baseline to zero
•
Append
: appends the baseline to the previous signal
•
Free
: lets the baseline "jump" to the current absolute value
TIP
You can perform the
AutoZero
command (automatic null balancing)
any time during data acquisition, that is, the absorption measured at the
time of the autozero is subtracted from any absorptions measured
thereafter. Always select a time for
AutoZero
at which no sample flows
through the cell.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...