8
•
Troubleshooting
Page 162
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Message and Code
Description and Remedial Action
Code 3116 Data transfer error from
detector to PC – check USB connection.
Check the USB connection. Use only the USB cables provided by
Thermo Fisher Scientific for the detector. Avoid CPU-intensive and
time-consuming operations on the PC during high-speed data
acquisition.
Code 3122
xx overtemperature - emergency standby
in yy seconds.
xx = Affected sensor (system (electronics) or lamp house)
yy = Number of seconds before detector is shut down
•
Turn off the detector and let it cool down.
•
Lower the environmental temperature.
Code 3123
xx overtemperature - emergency standby
after acquisition completed (force
shutdown in yy s).
xx = Affected sensor (system or lamp house)
yy = Number of seconds before detector is shut down
The given temperature sensor signals an overheating condition.
•
Wait for the acquisition to stop (max. 10 minutes).
•
Turn off the detector and let it cool down.
•
Lower the environmental temperature.
Code 3125 Wavelength calibration failed
– check and flush the flow cell.
•
Check whether the flow cell is installed properly.
•
).
•
Flush the flow cell with pure HPLC-grade water for several
minutes with a flow rate of > 1 mL/min.
•
The lamps should be running for at least 1 hour before you start
the wavelength calibration.
•
Check the UV lamp: If the UV lamp was newly installed, ensure
that it is installed properly. If the UV lamp is old (counting, for
example, more than 2000 operating hours), replace the UV lamp
(see
).
•
Repeat the wavelength calibration.
•
If the problem persists, contact Technical Support.
Code 3126 Slit width change failed –
check flow cell.
•
Change the slit width.
•
Check whether the flow cell is installed properly.
•
Remove the flow cell and check it for transparency. If it is not
transparent, clean the flow cell (see
).
•
Re-install the flow cell and try again.
Code 3127 Bad flow cell position.
•
Check whether the flow cell is installed properly.
•
Turn off the lamps, remove the flow cell from the detector.
Check whether the pins are dirty, bent or touch each other or
the optics. If the pins are bent or touch each other or the optics,
contact Technical Support. If the pins are dirty, turn off the
detector and clean them with isopropanol and a lint-free cloth.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...