Maintenance and Service
Page 136
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
6. Firmly squeeze the handle of the flow cell and carefully pull it from
the opening in the detector.
CAUTION—UV Radiation
The deuterium lamp emits UV radiation that is harmful to eyes and skin.
Do not look and do not reach into the flow cell opening while no flow
cell or diagnostic cell is installed and the UV lamp is turned on.
7.
If the flow cell is to be stored
Store the flow cell in its original packaging that it was shipped in.
8.
If a flow cell is to be installed
Install the flow cell (see
9. Install the flow cell cover.
NOTICE
The flow cell opening on the device is sensitive to dust and debris.
If no flow cell is installed to the device, close the flow cell opening with
the cover for the flow cell opening.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...