9
•
Specifications
Page 170
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
9.1
Performance Specifications
9.1.1
Detector
The device performance is specified as follows:
Type
Specification
Optical design
•
Single-beam, reverse-optics design with concave holographic grating
•
1024-element photodiode array
Light source
•
Deuterium lamp for UV and VIS wavelength ranges (30 W)
•
Tungsten lamp for VIS and NIR wavelength ranges (20 W) (standard
for VF detectors, optional for VC detectors)
Wavelength range
190 – 800 nm
Spectral bandwidth
0.6 nm pixel resolution (average); < 1 nm over complete range
Diode bunching (VC‑D11 and
VF‑D11 only)
1 – 100 nm, individually programmable for each signal channel and 3D
field
Wavelength accuracy
± 1 nm
Wavelength repeatability
± 0.1 nm
Wavelength calibration
Internal calibration with D-alpha line of the deuterium lamp
Wavelength validation
Internal validation with holmium-oxide glass filter
Signal channels
VC detectors:
8 signal channels
VF detector:
10 signal channels
Data collection rate
VC detectors:
Adjustable, 0.2 Hz – 125 Hz
VF detector:
Adjustable, 0.2 Hz – 250 Hz
Filter response times (in seconds)
VC detectors:
0 (no filter), 0.05, 0.1, 0.2, 0.5, 1,
2, 5, 10, 20
VF detector:
0 (no filter), 0.01, 0.02, 0.05, 0.1,
0.2, 0.5, 1, 2, 5, 10, 20
Spectra scan
VC‑D11:
3D field with full
spectral range 125 Hz
VC‑D12:
Not available
VF‑D11:
3D field with full
spectral range 250 Hz
Noise
<± 6 µAU at 254 nm ASTM
<± 12 µAU at 520 nm ASTM
Noise reference conditions:
•
Wavelength: as listed above; bandwidth 4 nm; reference wavelength:
Off; slit width: Wide
•
Time constant: 2 sec (response time = 2.2 x time constant as proposed
in ASTM). For details, see below this table.
•
Flow cell: 10 mm standard stainless steel flow cell
•
Flow: 1 mL/min LC/MS-grade water from highly efficient water
purification system; column installed in flow path before flow cell;
pump pressure: >7 MPa
•
Temperature conditions: See below this table.
Содержание Vanquish VC-D11
Страница 4: ...Contacting Us Page 4 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 10: ...Contents Page 10 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 42: ...3 Device Overview Page 42 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 48: ...4 Unpacking Page 48 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 168: ...8 Troubleshooting Page 168 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 186: ...11 Appendix Page 186 Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual...
Страница 191: ...Index Diode Array Multiple Wavelength Detectors VC D11 VC D12 VF D11 Operating Manual Page 191...
Страница 192: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...