Changing the Column
Connecting the Column to the Transfer Line
26
TSQ 8000 Evo User Guide
Thermo Scientific
To connect the column using the regular transfer line nut
1. Lower the oven temperature and allow it to cool.
2. If the TSQ 8000 Evo system is running, shut down and vent it. See the instrument’s
hardware manual for instructions.
3. Unwind about one turn of the column from the column outlet end.
4. Wipe approximately 300 mm (12 in.) of the column with a tissue soaked in methanol.
5. Choose an appropriate ferrule for the outer diameter of your column.
6. Insert the column through the transfer line nut and ferrule, entering through the tapered
end of the ferrule. Wipe the column again with a tissue soaked in methanol.
Figure 24.
Transfer Line Nut and SilTite Ferrule Orientation
Note
For best results, we recommend you use the spring loaded transfer line nut. See
connect the column using the spring loaded transfer line nut”
CAUTION BURN HAZARD:
The injector, detectors, oven, and transfer line may be hot.
Allow them to cool before touching them.
Note
Wear clean, lint- and powder-free gloves when you handle the column and
transfer line ferrule.
Note
If the maximum oven temperature in your method is
≥
290 °C (554 °F),
Thermo Fisher Scientific recommends using a spring loaded transfer line nut with a
graphic Vespel ferrule or a SilTite™ nut and ferrule. By cycling the oven at and above
this temperature, expansion and contraction of the graphite Vespel material can cause
leaks in the transfer line.
Transfer Line Nut
SilTite Ferrule
Flat on the
Transferline Nut