Changing the Column
Connecting the Column to the Transfer Line
Thermo Scientific
TSQ 8000 Evo User Guide
27
7. Insert the column into the measuring tool, which is in the MS Toolkit (See
that it is even with the lines at the end of the column.
positioning of the column in the tool for accurate measuring.
8. Use a scoring wafer to score and break the column. Use a magnifying glass to check for an
even, flat cut. Repeat if necessary.
9. Use a 5/16 in. wrench to hold the column measuring tool steady.
Figure 25.
Column Measuring Tool
10. While holding the column measuring tool steady, tighten the transfer line nut with a 1/4”
wrench until the column just stops moving in the ferrule.
11. Turn the transfer line nut 1 flat backward so the column is able to move in the ferrule
with slight resistance.
12. Line up the outlet of the column with the arrows on the end of the column measuring
tool.
Figure 26.
Lining Up the Column in the Column Measuring Tool
13. Place a septum with a notch cut into it behind the transfer line nut. The septum marks
the place on the column where it should exit the nut.
Column Outlet