7
Troubleshooting
Investigating Peak Issues
122
TSQ 8000 Evo User Guide
Thermo Scientific
Tailing
Sample peaks
Column degradation is
causing activity
Inject a test mixture and evaluate the
column. Replace column if necessary.
Liner is dirty
Clean or replace the liner.
Ion source is dirty
Clean the ion source and tune the
instrument.
Glass wool or inlet liner is
causing activity
Replace wool with fresh silanized wool
and install a clean inlet liner.
Inlet temperature is too low Increase the temperature of the inlet.
Column connections are
poor or obstructed
Reconnect the column inlet.
Stationary phase is not
appropriate for your target
compounds
Replace the column and choose a more
appropriate phase for your analysis.
Final hold oven
temperature is too low
crease the column/oven temperature. Do
not exceed the recommended maximum
temperature for the stationary phase.
Transferline temperature is
too low
If tailing occurs on late eluting
compounds, it is likely the source or
transferline temperature is too low.
Source temperature is too
low
If tailing occurs on late eluting
compounds, it is likely the source or
transferline temperature is too low.
Poor column
characterization
information about checking for leaks
and column flow.
Table 4.
Troubleshooting Peak Issues in Your Data, continued (Sheet 2 of 3)
Behavior
Characteristic
Cause
Remedy
Содержание TSQ 8000
Страница 1: ...TSQ 8000 Evo Mass Spectrometer User Guide 1R120586 0002 Revision C December 2015 ...
Страница 22: ......
Страница 148: ......