15. Select the
Filament 2
check box to enable the maintenance reminder. Then set the
number of days after which you want to be reminded to check filament 2. In a leak-free
system, expect the filament to last between 30-360 days, depending on usage.
16. Select the
Ion source
check box to enable the maintenance reminder. Then set the
number of days after which you want to be reminded to check the ion source. The time
between cleaning depends very strongly on your analysis. You will have to determine the
correct length of time between source cleanings.
17. Select the
Multiplier
check box to enable the maintenance reminder. Then set the
number of days after which you want to be reminded to check the electron multiplier.
18. Click
OK
to return to the main Instrument Configuration window. See
Figure 12.
Main Instrument Configuration Window
19. Select the TRACE 1300 or TRACE 1310 icon to configure the GC handshaking
parameters.
20. Under the
General
tab, set the handshaking parameters as shown in
.