Creating a Method
Creating an Autosampler Method
48
TSQ 8000 Evo User Guide
Thermo Scientific
e.
(Optional) Cooled/Heated Trays—Enter ID information for your primary and, if
used, secondary trays.
4. Click
Create New Method
.
5. Use these instructions to configure the GC Liquids general method settings (see
Figure 48.
General GC Liquids Settings on the TriPlus RSH Sampling System
a. Injector Type—Choose Single or Double for each injector.
b. Injection Mode—Choose Basic, Enrichment, Enrichment Needle Solvent Option,
Internal Standard Double, Internal Standard Post, Needle Solvent Wash, Solvent
Flush Double, or Solvent Flush Post.
c.
Rapid Mode—Choose Disable, After Sample Rinse, After Bubble Elimination, or
After Sample Asp. in Home.
d. Syringe Type—Enter your syringe volume in μL. Enter the syringe needle length in
mm.
e.
Sample
Volume—Enter the sample volume to be injected into the GC. Typical values
are between 0.5 and 5 μL.
f.
Plunger
Strokes—Select the number of plunger strokes to use when drawing up the
sample. Air bubbles in the syringe change the amount of sample injected, which can
cause signal variation in different runs. To prevent this from occurring, increase the