Tuning the Mass Spectrometer
Thermo Scientific
TSQ 8000 Evo User Guide
33
Figure 32.
Tuning Categories
3. Select the tune type you want to use:
Figure 33.
Tune Types
a.
EI Standard Tune
—Provides EI tuning and is used to completely retune the system.
It takes the longest amount of time to run, but it has the advantage of reoptimizing
nearly all the parameters affecting the signal. This type of tune performs a mass
calibration, tunes the lenses and resolution, and performs a leak check. It adjusts the
detector sensitivity to generate a
m/z
219 ion with an intensity of 20,000,000 counts.
Unless your SOP requires it, this is not the best tune to use on a daily basis because of
the length of time required to run it.