4
Creating a Method
Creating a MS Instrument Method
Thermo Scientific
TSQ 8000 Evo User Guide
61
viii. Use the
CI Gas Flow
column to set the flow rate of your reagent gas. Click the
Apply to All Groups Button
at the top of the method editor. The single value at
the top of the method is sent when initializing the MS with the method.
Figure 58.
Acquisition-General CI Settings
ix. The
MS/MS Gain Multiplier
column only applies to groups with MS/MS
events. The MS/MS gain multiplier only applies if the
Use Last Tuned Detector
Gain
column is selected. The default MS/MS Gain Multiplier is 7 and is
intended to give similar signal levels as a SIM scan on the precursor ion. If you
enter a specific gain, the MS will use that gain for the SRM scan without the
multiplier.
5. The following settings apply to
Acquisition-Timed
methods only.