TSQ 8000 Evo User Guide
115
7
Troubleshooting
This section contains information to help you diagnose problems within your data.
Sometimes, your experience as a scientist will enable you to look at your data and detect that
something is wrong either with the instrument or your sample. This chapter describes the
most common indications of a problem with a baseline, peak or result.
If there is an issue with the hardware, see the troubleshooting section of the hardware manual.
You can begin troubleshooting by running a tune with diagnostics on the instrument. If you
have good ion intensities, good peak shapes, and no air leak, you might want to look first at
the GC, autosampler, or carrier gas.
If you have air leaks, locate and address them. Pay particular attention to the transfer line
ferrule, vent valve knob, front panel, gas quality, carrier gas vacuum compensation, and
vacuum interlock on the mass spectrometer, as well as the inlet on the GC.
If your intensities are too low, make sure carrier vacuum compensation is turned on.
Contents
•
•
How to Know When Your System Needs Maintenance
•
•
•
CAUTION ELECTRICAL SHOCK HAZARD:
When troubleshooting any issue that requires
removing a cover on the instrument, power-off and vent the instrument to avoid possible
injury.