7
Troubleshooting
Investigating Results Issues
124
TSQ 8000 Evo User Guide
Thermo Scientific
Investigating Results Issues
Behavior
Characteristic
Cause
Remedy
Low
Reproducibility of
Peaks Area
General
Detector gain is set too low Retune the gain.
Increase the electron multiplier voltage.
Increase the target ion count.
Concentration is not
compatible with the
dynamic range of the
detection system
Verify that the sample concentration is
suitable for the system.
Chromatogram and
spectrum are blank
Make sure the tune file is correct.
Verify that the Busy light is on during
acquisition.
Make sure the filament is not burned
out.
Injection technique is not
appropriate
Use a different injection technique.
Injection parameters are
not appropriate
Verify the injection temperature and
flow rates.
Sample injection technique
is not reproducible
Evaluate the sample preparation
sequences.
Compare the results with a series of
standard injections.
Syringe or septum is
leaking
Check and replace the syringe and/or
septum at regular intervals.
Injection port is leaking
Check the column connections.
Run a leak check.
Injection technique is not
suitable
Carefully meter the injected amount.
Use a clean, good-quality syringe.
Ion source is dirty
Clean the ion source and retune the
instrument.
Split flow or ratio control is
inadequate
Monitor the flow.
Replace the in-line filter.