Creating a Method
Creating a MS Instrument Method
58
TSQ 8000 Evo User Guide
Thermo Scientific
x. Use the
Ion Polarity
column if you are using CI mode to tell the instrument to
generate positive or negative ions. Only use this column if you are using CI
mode. In EI mode, this column should always be set to
Positive
.
xi. Use the
Data Type
column to select whether you want to collect
Profile
,
Centroid
, or
Nominal
mass spectra. Centroided mass spectra are the most
common because they are used by most of the libraries and provide the smallest
data files. Profile mass spectra provide detailed mass spectral peaks, which result
in large data files. Profile mass spectra are not available for SRM scans. when you
want to perform fast scanning (up to 20,000 amu/s), select
Nominal
from the
drop-down list under
Data Type.
xii. Use the
Precursor (Q1) Resolution
column to open the peak width of the
precursor mass for SRM scans. The options are
Normal
,
Wide (1.5 at 50%)
, or
Widest (2.5 at 50%)
. The default setting is Normal, which uses the tuned
resolution values from the tune file specified in the
Tune File Name
column.
Selecting Wide or Widest increases the precursor peak width to 1.5 and 2.5 amu
(at 50%) respectively. Opening the precursor peak width will increase the ion
signal into the collision cell and may improve the sensitivity of your method;
however, it may also increase the background noise level.
xiii. Use the
Product (Q3) Resolution
column to open the peak width of the
precursor mass for SRM scans. The options are
Normal
,
Wide (1.5 at 50%)
, or
Widest (2.5 at 50%)
. The default setting is Normal, which uses the tuned
resolution values from the tune file specified in the
Tune File Name
column.
Selecting Wide or Widest increases the precursor peak width to 1.5 and 2.5 amu
(at 50%) respectively.
c.
In the
Groups
pane on the right, review the information in each row. See
As you create scans in the Scans pane, information in the groups pane is
automatically displayed.
Note
By default not all columns are displayed in the Groups pane. Right-click a
heading to display them. You can also reorganize the columns by clicking the
heading of a column and dragging it to the left or right.