5
Optimizing Your Method
Modifying an Automatic Tune
Thermo Scientific
TSQ 8000 Evo User Guide
97
b.
Description
—Enter details or notes about your tune type.
c.
Categories
—This is an optional field that allows you to display a subset of tunes
under Auto Tune on the dashboard. This is convenient for separating your most
frequently used tunes.
d.
Type
—Select the
Tune and Diagnostics
option to run a tune with diagnostics or
select the
Diagnostics Only
option if you are creating a diagnostics test.
e.
Output Tune Filename Prefix
—Enter a prefix to be added to the title of your tune
report. It is a good idea to use a descriptive prefix for your tune type. For example, if
you are always running BFB reports, you could enter BFB here to distinguish it from
other reports you are generating.
f.
Starting Tune File
—Select a tune file that your tune type will be based on:
• Select
Factory
to use a default factory-made file on the instrument that can be
used to begin tuning an instrument with a clean ion source.
• Select
Last Saved
to use a tune file saved on the instrument by the most recent
successful automatic tune.
• Select a specific tune file if you have a reliable tune file you want to use as a
starting point for a new tune file.
g.
Perform Mass Calibration
—Select this to enable the system to recalibrate all of the
masses during a tune.
h.
Check Mass Calibration
—Select this to enable the system to confirm that your mass
calibration is correct rather than performing a mass calibration.
4. Select the
Ion Source
tab to configure the ion source as follows:
Note
Create the new tune type name now. You will not be able to edit the tune
type name later. Also, you will not be able to save new parameters to a built-in
tune.
Note
If you select a tune file with a prefix-only name rather than a tune file with
a date-time stamp, the most recent tune file of with that prefix name will be
loaded at the start of each tune.
Note
If Tune and Diagnostics is selected, you can select only one Mass
Calibration option.