Troubleshooting
Error Messages
42
Surveyor RI Plus Detector Hardware Manual
Thermo Scientific
Table 8.
Pop up error messages and their solutions
Error message
Remedy
Initialize Device failed.
The detector is not connected to the
data system. Shutdown and restart the
instruments.
Direct Control failed.
The device is not available. Shutdown
and restart the instruments.
Set Config Settings failed.
Shutdown and restart the instruments.
Terminate Device failed.
Shutdown and restart the instruments.
System Is Ready For Run but failed.
Shutdown and restart the instruments.
Initialize Device For Run failed.
Shutdown and restart the instruments.
Lost connection to Surveyor RI on stack.
Check the instruments connections.
Shutdown and restart the instruments.
Acquisition Is Complete but failed.
Restart the run.
Start Run failed.
Shutdown and restart the instruments.
Abort Run failed.
Restart the run.
Table 9.
Error messages that appear in the RI Status page and their solutions
Error message
Remedy
Potential data error during acquisition
Restart the run.
RI leaking
Check all of the fittings for leaks.
Null glass movement cannot detect home
Purge the detector. Flush the flowcell
with fresh mobile phase.
Heater error
Malfunction with heater circuit. Call
your Thermo Fisher Scientific service
representative for assistance.
Optical balance movement cannot detect home
Call your Thermo Fisher Scientific
service representative for assistance.
RAM error
Instrument memory failure. Call your
Thermo Fisher Scientific service
representative for assistance.
Light intensity too low
Purge the detector. Flush the flowcell
with fresh mobile phase.
Instrument connection loss
Check Ethernet connection and start the
software. Check the module
configuration in the software.
Extend Run failed.