Troubleshooting
Detector-Related Problems
Thermo Scientific
Surveyor RI Plus Detector Hardware Manual
39
3. Excessive baseline
drift Drift.
a. Flowcell contaminated.
a. Flush the flowcell with
cleaning solvents as
described in
leaks.
b. Mobile phase
contamination.
b. Replace with fresh mobile
phase made with
high-purity solvents.
c.
Material bleeding from
column.
c.
Clean or replace the
column.
d. Leaks in system or flowcell. d. Check all of the fittings for
leaks. Call a Thermo Fisher
Scientific Service
Representative to replace
the flowcell.
e.
Tiny bubble trapped in
flowcell.
e.
Increase the flow rate until
the bubble is removed.
Connect a backpressure
device to the flowcell outlet
(check backpressure rating
to avoid rupturing
flowcell).
f.
Large temperature
fluctuations.
f.
Remove system from drafts.
Thermostatically control
column temperature.
Table 7.
Troubleshooting detector-related problems (Sheet 3 of 5)
Symptom
Cause
Remedy