3
Operation
Calibrating the Detector
30
Surveyor RI Plus Detector Hardware Manual
Thermo Scientific
2. Connect a multimeter to the pin connector adapter. Ensure that you connect the positive
connector of the multimeter to the positive pin of the adapter. Ensure that you connect
the negative connector of the multimeter to the negative pin of the adapter.
3. Set the multimeter to mV.
4. Click the
RI Diagnostic
tab.
5. In the Calibration list, select
Recorder Output Adjustment
. See
6. Click
Go
.
7. Adjust the recorder settings by clicking
Up
or
Down
as shown in
. Use
Up
or
Down
to increase or decrease the value on the multimeter. A “Gain” message appears in
the Progress area.
Figure 20.
Adjusting the recorder settings
8. Click
Done
when you have completed the adjustment.
9. Disconnect the multimeter from the pin connector adapter.
10. Disconnect the pin connector adapter from pins 9 and 10, located on the back of the
detector.
Calibration list
Adjustment settings
buttons