Dionex Integral Systems Operator’s Manual
128
Doc. 065250-06 03/2020
5. Carefully remove the pump head from the SP by pulling the head straight off
and away from the SP panel.
6. Place the head on a clean work surface.
7. Hold the shaft of the piston (near the base), tilt the piston slightly, and pull the
piston away from the pump. The piston does not come off as part of the pump
head assembly because it is captured by a magnetic retention system.
To install the new backup seal:
1. Lift the spacer off the pump head.
2. Remove the backup washer from the spacer, using
one of the following methods:
•
With the backup washer facing up, cup the
spacer in your hand and blow clean laboratory
air at the spacer to dislodge the washer.
•
If air is not available, create a removal tool by
using a pair of pliers to sharply bend the end of a
paper clip (see the following photo). Carefully
Lateral motion while disengaging the pump head from the piston may
break the piston.
Un mouvement latéral pendant la séparation de la tête et du piston
peut casser le piston.
Vermeiden Sie Seitwärtsbewegungen, wenn Sie den Pumpenkopf vom
Kolben lösen. Andernfalls kann der Kolben brechen.
Backup
Washer
Spacer
Содержание Dionex
Страница 1: ...Dionex Integral Process Analytical Systems Operator s Manual Document No 065250 Revision 06 March 2020...
Страница 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Страница 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Страница 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Страница 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Страница 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Страница 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Страница 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Страница 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Страница 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Страница 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Страница 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...