3 • SP Sample Preparer Description
Doc. 065250-06 03/2020
49
3.8.1
Relay Output Control
The relay outputs are normally open (NO). This means that the relay is
open when it is off (not energized) and closed when it is on (energized).
The relays can be programmed to switch any low-voltage device.
Switched current must be no more than 2 A at 24 Vdc. For connection
instructions, see
.
To control relay outputs, use any of the following methods.
•
To change the relay's state immediately (direct control), go to the SP
ePanel in Chromeleon 7.
•
To program the relay’s state change in an instrument method, add the
control commands in the Script Editor view in the Chromeleon 7
Instrument Method Editor.
Notes:
•
If a relay output is configured to change states with an alarm
event (see
), do not program the relay output to
control devices.
•
Always include a command that opens (turns off) the relay at the
beginning of the instrument method. Include the command in all
methods used with the system.
•
To configure the relay's state change as a function of an alarm
condition or a TTL input state change, see
Responses (Actions) in Chromeleon 7 PA” on page 50
).
3.8.2
TTL Input Control
The eight TTL inputs allow an external device to trigger one or more of
the following responses (actions):
•
Shut down the system
•
Put the system in standby
•
Bypass an injection from a specified sample stream
•
Turn on an SP panel light
•
Turn on an SP relay
Содержание Dionex
Страница 1: ...Dionex Integral Process Analytical Systems Operator s Manual Document No 065250 Revision 06 March 2020...
Страница 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Страница 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Страница 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Страница 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Страница 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Страница 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Страница 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Страница 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Страница 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Страница 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Страница 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...