9 • Troubleshooting
Doc. 065250-06 03/2020
105
If a sample preparation command is assigned a specific retention time,
position the command outside the sample preparation command block.
This error occurs if the instrument method includes a
DelaySP
command, but
the command is not inside the sample preparation command block. See
for more information.
To troubleshoot:
Move the command inside the sample preparation command block.
–or–
Delete the sample preparation command from the method.
This error may indicate a problem in the dilution pump electronics.
To troubleshoot:
1. Turn off the pump flow.
2. Turn off the SP power briefly by pressing the SP power switch.
3. Turn on the SP power again and restart the pump flow.
4. If the problem persists, contact Technical Support for Dionex products for
assistance.
DelaySP command must occur within BeginSPSamplePrep/
EndSPSamplePrep command block
Dilution pump driver is over current
Содержание Dionex
Страница 1: ...Dionex Integral Process Analytical Systems Operator s Manual Document No 065250 Revision 06 March 2020...
Страница 12: ...Dionex Integral Systems Operator s Manual x Doc 065250 06 03 2020...
Страница 17: ...1 Introduction Doc 065250 06 03 2020 5 Figure 1 2 Example Analyzer Configuration...
Страница 28: ...Dionex Integral Systems Operator s Manual 16 Doc 065250 06 03 2020...
Страница 72: ...Dionex Integral Systems Operator s Manual 60 Doc 065250 06 03 2020...
Страница 92: ...Dionex Integral Systems Operator s Manual 80 Doc 065250 06 03 2020...
Страница 96: ...Dionex Integral Systems Operator s Manual 84 Doc 065250 06 03 2020...
Страница 154: ...Dionex Integral Systems Operator s Manual 142 Doc 065250 06 03 2020...
Страница 160: ...Dionex Integral Systems Operator s Manual 148 Doc 065250 06 03 2020...
Страница 202: ...Dionex Integral Systems Operator s Manual 190 Doc 065250 06 03 2020...
Страница 214: ...Dionex Integral Systems Operator s Manual 202 Doc 065250 06 03 2020...
Страница 228: ...Dionex Integral Systems Operator s Manual Index 10 Doc 065250 06 03 2020...