Profiles when the SAE module is enabled
After the SAE module is enabled on the instrument, the local instrument profiles and the Connect
cloud-based platform profiles will not be available.
An account from the SAE Administrator Console must be used to sign in to the instrument when the
SAE module is enabled. A local instrument administrator can sign in to the instrument to perform limited
administrator functions.
The local instrument profiles and the Connect cloud-based platform profiles are available if the SAE
module is disabled.
SAE-enabled system components
The following system components can be used with SAE functions enabled:
• QuantStudio
™
Design and Analysis Software v2
• QuantStudio
™
7 Pro Real-Time PCR System
• QuantStudio
™
7 Pro Real-Time PCR System plate file—SAE-enabled plate files are created in the
QuantStudio
™
Design and Analysis Software v2 with SAE enabled.
– If SAE is enabled in the QuantStudio
™
Design and Analysis Software v2, you can only create
plate files for the QuantStudio
™
7 Pro Real-Time PCR System.
– If SAE is enabled on the QuantStudio
™
7 Pro Real-Time PCR System, you cannot create or edit
a plate file from the instrument touchscreen.
• QuantStudio
™
7 Pro Real-Time PCR System data file—SAE-enabled data files are created on the
QuantStudio
™
7 Pro Real-Time PCR System with SAE enabled.
We recommend enabling SAE for all system components (see “Enable SAE functions”
one or more of the components have conflicting SAE status, some functions may not be available (see
“Compatibility between SAE-enabled and SAE-disabled components”
Chapter 9
Use the instrument with the Security, Auditing, and E
‑
signature (SAE) v2.0 module
Profiles when the SAE module is enabled
9
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide
141