Figure 10 Home screen during a run
1
Arrow
2
Actions menu
Edit a plate file during a run
The targets and samples can be added or edited after a run has started. The method cannot be edited
after a run has started.
1.
In the home screen during a run, tap
(Actions)
.
.
2.
Tap
Set up
.
3.
In the
Plate
tab, tap a well or multiple wells.
Note:
Use the plate view to edit or assign a well (see “Options to view a plate layout”
4.
Tap the
Sample Information
pane, then tap
Samples
.
Note:
If no wells are selected, the pane is labeled
Legend
.
5.
Enter a sample name, then select a sample type from the dropdown list.
• Unknown
• Standard
• No template control (NTC)
6.
Tap
Target/SNP
, then select
Target
or
Assay
from the dropdown list.
7.
Tap a row in the table, then tap
Close
4
Done
.
Chapter 5
Options for a run
Edit a plate file during a run
5
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide
71