Pause an instrument run
Pause a run to add reagents to the plate or to edit the number of cycles.
1.
In the home screen during a run, tap
(Actions)
.
.
2.
Tap
Pause run
, then enter a pause temperature.
3.
(Optional)
Tap
Edit
to change the number of cycles.
4.
(Optional)
Tap
(Eject)
to open the drawer, then tap
(Eject)
to close the drawer.
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the plate temperature
can reach 100°C. If you want to access the plate during a run pause, enter room temperature as
the pause temperature and allow the plate to cool to room temperature before handling.
Stop an instrument run
CAUTION!
PHYSICAL INJURY HAZARD.
During instrument operation, the plate temperature can
reach 100°C. If you want to access the plate during a run pause, enter room temperature as the
pause temperature and allow the plate to cool to room temperature before handling.