1
Figure 8 VeriFlex
™
Zones and Auto Delta settings
1
VeriFlex
™
Zones
is selected
2.
Tap the temperature field in the appropriate VeriFlex
™
Zone.
3.
Enter a temperature, then tap
Enter
.
The temperatures in adjacent zones must have a difference of ≤5°C.
4.
Tap
Done
.
A stage with VeriFlex
™
Zones is denoted with a icon.
5.
In the
Method
tab, tap
4
Remove
4
Done
to remove the VeriFlex
™
Zones.
Note:
The VeriFlex
™
Zones setting is removed from the
Method
tab.
1
2
1
Indicates that this is the
Method
tab where the VeriFlex
™
Zones settings can be removed
2
Indicates that VeriFlex
™
Zones are assigned. Tap to remove the VeriFlex
™
Zones.
Chapter 4
Edit a plate before starting a run
Options to edit the method of a plate file
4
60
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide