Operation
Calibration
Thermo Scientific
48iQ Instruction Manual
3-17
The Advanced Calibration screen provides several additional ways to
calibrate the instrument and view the calibration history. See Chapter 4
“
Calibration
” for further instructions on how to run a calibration.
The following screens show the advanced screens in single range mode and
dual or auto range mode. (The only difference between the screens, are the
words “High” and “Low”.) For more information about range modes, see
“
Range Mode Selection
” on page
3-76
.
Home Screen>Calibration>Advanced Calibration (single range mode
with O
2
option)
Home Screen>Calibration>Advanced Calibration (dual or auto range
mode with O
2
option)
The Advanced Calibration screens contain the following information:
●
Manual Calibration:
The user manually adjusts the background or span
coefficient.
●
Multipoint Calibration:
Up to three gas concentrations (cal-points) may
be calibrated. Three cal-points will give the most accurate readings over
the entire range.
Advanced
Calibration
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......