Introduction
Specifications
Thermo Scientific
48iQ Instruction Manual
1-5
Table 1–1
lists the specifications for the 48iQ.
Table 1–1.
48iQ Specifications
Range
0-10000 (ppm or mg/m
3
)
Zero Noise
0.02 ppm RMS (30 second averaging time)
Detection Limit
0.04 ppm (30 second averaging time)
Zero Drift
<0.1 ppm (24 hour)
Span Drift
±
0.5% reading (24 hour)
Response Time
60 seconds (30 second averaging time)
Linearity
±
1% full-scale ≤ 1000 ppm
±
2.5% full-scale > 1000 ppm
Flow Rate
1.0 SLPM (1 atm inlet pressure)
Operating Temperature Range
5–
45
°
C (may be safely operated over the range of 0–
45
°
C)
Power Requirements
100–240 VAC 50/60 Hz
275 watts
Physical Dimensions
24 in (D) x 16.75 in (W) x 8.72 in (H) [609 mm (D) 425.45 mm
(W) x 221.48 mm (H)]
Weight 34.3
lbs
Analog I/O
4 Isolated Voltage Inputs 0–10 V
6 Isolated Analog Voltages Outputs, with 4 selectable
ranges
6 Isolated Analog Current Outputs, with 2 selectable ranges
Digital I/O
16 Digital Inputs (TTL)
8 Solenoid Driver Outputs
10 Digital Reed Relay Contact Outputs
Serial Ports
1 RS-232/485 port
1 RS-485 External Accessory port
Other Ports
3 Full Speed USB ports (one in front, two in rear)
1 Gigabit Ethernet port
Communication Protocols
MODBUS, Streaming, Bayern Hessen
Approvals and Certifications
CE, TUV-SUD Safety, EPA
Specifications
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......