Servicing
Optional DMC Oxygen Sensor
Thermo Scientific
48iQ Instruction Manual
7-51
Use the following procedure to remove the oxygen sensor from the
instrument case. It is easier to do the following by removing the oxygen
sensor first then remove and replace the following as necessary:
●
Oxygen Sensor removal
●
Oxygen Sensor board replacement
●
Oxygen Sensor capillary replacement
●
Oxygen Sensor replacement
Use the following procedure to remove and replace the oxygen sensor.
Equipment required:
Phillips drive, #2
1.
Turn instrument OFF, unplug power cord, and remove the cover
(
Figure 2–1
).
2.
Unplug electrical, unplug DMC cable to oxygen sensor board DMC
(J4).
3.
Disconnect 2X plumbing.
4.
Using a #2 Phillips drive, unfasten captive hardware.
Optional DMC
Oxygen Sensor
Oxygen Sensor
Removal
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......