Operation
Calibration
Thermo Scientific
48iQ Instruction Manual
3-29
The 2-Point O
2
Calibration screen is used to calibrate the O2 sensor using
2 different gas concentrations. Both points need to be calibrated for a
successful calibration. It is suggested that the two O
2
concentrations used
for this calibration be 0% and 20.9%. This button appears if the O
2
Sensor
option is selected in the Configuration screen.
Home Screen>Calibration>Advanced Calibration>2-Point O
2
Calibration>
Home Screen>Calibration>Advanced Calibration>2-Point O
2
Calibration>Continue to 2-Point O
2
Calibration
The 2-Point O
2
Calibration screen contains the following information:
●
2-Point O
2
Calibration Point 1:
First cal point used in the 2 point O
2
calibration. Defaults to 0.000%.
●
2-Point O
2
Calibration Point 2:
Second cal point used in the 2 point O
2
calibration. Defaults to 20.900%.
2-Point O
2
Calibration
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......