Operation
Data
3-38
48iQ Instruction Manual
Thermo Scientific
The Data Logging Setup screen allows the user to select data to be stored
and how it is stored.
Home Screen>Data>Advanced>Data Logging Setup
The Data Logging Setup screen contains the following information:
●
Select Data Logging Variables
: User selects instrument variables to log.
●
Period
: User selects how often data is collected by setting the duration
between logged data.
●
Data Treatment
: Toggles between Average, Current, Minimum and
Maximum. When set to average, the average value during the period
will be recorded. When set to current, the latest data will be recorded.
When set to minimum or maximum, the minimum or maximum value
during the period will be recorded.
●
Erase Data Log Records
: Allows the user to erase all values in the data log
and updates the data logging table.
Data Logging Setup
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......