Servicing
DMC Optical Bench
Thermo Scientific
48iQ Instruction Manual
7-41
Use the following procedure to remove the optical bench or replace heater
assembly.
Equipment required:
Hex drive, 5/32
Phillips drive, #1
Allow to cool before handling bottom of optical bench.
▲
1.
Unplug electrical connections, J5 heater, J1 preamp, bottom
connection J8 motor, and top connection J4 opt.
2.
Using a 5/32 hex drive, unfasten four #10-32 socket cap head screws
and lockwashers.
Figure 7–44.
Bench Removal pt 1
Optical
Bench/Heater
Removal
Icon Here