Maintenance
Cleaning the Optics
5-6
48iQ Instruction Manual
Thermo Scientific
Best results are obtained when the optics are cleaned prior to calibration.
The cleanliness of the mirrors should be checked any time the intensity is
below 200,000 Hz, since one cause of low output is light attenuation due
to dirt on the mirrors.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component. For more
information about appropriate safety precautions, see the “
Servicing
”
chapter.
▲
Use the following procedure to clean the mirrors.
Equipment Required:
Hex drive, 9/64
1.
Turn off power and disconnect power line.
2.
Using 9/64 hex drive, remove the field mirror by removing the four #8-
32 socket cap head screws holding it to the main bench.
Figure 5–3.
Removing the Field Mirror
Cleaning the
Optics
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......