Optional Equipment
Connecting External Devices
Thermo Scientific
48iQ Instruction Manual
9-3
The RS-485 external accessory port uses a 15-pin serial connector for
communication with external smart devices.
Figure 9–3.
RS-485 External Accessory Port
Table 9–2.
RS-485 External Accessory Port Terminal Assignment
Terminal Number
Signal Name
1 EXT_RS485_RX_N
2 EXT_RS485_RX_P
3 +5V
(Fused
@0.4A)
4 +5V
(Fused
@0.4A)
5 +5V
(Fused
@0.4A)
6 GND
7 GND
8 GND
9 EXT_RS485_TX_N
10 EXT_RS485_TX_P
11 +24V
(Fused
@0.4A)
12 +24V
(Fused
@0.4A)
13 +24V
(Fused
@0.4A)
14 +24V
(Fused
@0.4A)
15 +24V
(Fused
@0.4A)
RS-485 External
Accessory Port
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......