Operation
Settings
3-86
48iQ Instruction Manual
Thermo Scientific
When O
2
correction is enabled, the following equation is applied to the
concentration value to generate a corrected value. The corrected value may
be output on the analog outputs, stored in logging memory, or output as
streaming data. The corrected values are not displayed on the Home Screen
on the front panel, only the non-corrected values.
2
9
.
20
2
9
.
20
*
2
@
O
O
C
O
C
meas
corr
corr
comp
Abbreviations: C is the measured concentration, O2
corr
is the oxygen
concentration that C is corrected to (this value is set by the user in the
Source O
2
Concentration Setpoint screen and defaults to 15%). O2
meas
is
the measured O
2
concentration, which is clamped at 20.8% to prevent
divide-by-zero errors. C
comp
is the corrected concentration that may be used
for analog outputs or datalogging.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Compensation>O
2
Correction (single range mode)
O
2
Correction
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......