Operation
Calibration
3-30
48iQ Instruction Manual
Thermo Scientific
The Point 1 screen allows the user to view and set the selected calibration
point. The point 2 screen functions the same way. Therefore, the following
example of point 1 screen applies to point 2 as well.
Home Screen>Calibration>Advanced Calibration>2-Point O
2
Calibration>Continue to 2-Point O
2
Calibration>2-Point O
2
Calibration Point 1
The 2-Point O
2
Calibration Point 1 screen contains the following
information:
●
Edit O
2
Span Concentration:
User enters O
2
span concentration.
●
Current O
2
Concentration:
Read only. After selecting the value on the
Edit O
2
Concentration button and after pressing the Calibrate O
2
button, the Point 1 Concentration value is updated.
●
Calibrate O
2
Point 1:
Calibrates point 1 coefficient.
Note
Ensure that the sensor has been sampling the gas for at least 1 minute
before pressing the Calibrate O
2
Point 1 button.
▲
Point 1
2
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......