Operation
Settings
Thermo Scientific
48iQ Instruction Manual
3-121
4.
The Download Data to USB Drive screen will display. Select from
various options to download.
5.
The instrument will display a “downloading data” message and begin
transferring data to the USB drive.
Note
Do not remove the USB drive from the instrument while the data
is downloading
.
▲
6.
When the data download is complete, the instrument will display a
“Success!” message and display the file name as it is stored on the USB
flash drive. (The file name format is the instrument serial number,
name of download, followed by a date/time stamp.) Remove the USB
flash drive and select the OK button to continue.
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......