Calibration
Calibration
Thermo Scientific
48iQ Instruction Manual
4-9
Use the following procedure to set the CO reading to zero.
1.
Allow sufficient time for the instrument to warm up and stabilize.
2.
Adjust the dilution system of
Figure 4–1
so that zero air alone is
present in the manifold.
Since not all flow controllers have a positive shut off, it might be
necessary to disconnect the CO input line and cap it.
3.
Allow the instrument to sample zero air until a stable reading is
obtained.
4.
From the Home Screen, press Calibration>
Calibrate Background
.
The Target Concentration button will read 0.000. The Calculated
Background button will display the background needed to make the
current CO concentration go to zero.
5.
Press
Calibrate
to set the CO reading to zero and to save the new
background.
Calibrate
Background
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......