System Description
Oxygen Sensor (optional)
Thermo Scientific
48iQ Instruction Manual
8-9
●
Peripherals Support Board
●
Optional I/O (Communications, Analog, and Digital)
●
Optional Oxygen Sensor
The firmware contains the active controls for their application, as well as
self-identification and configuration for “plug and play” style operation.
Each are associated with specific registers of two types:
●
Modbus registers that are communicated from each microprocessor to
the System Controller Board (SCB) via internal RS-485
●
SNMP registers that are maintained in the software and SCB for health
and data processing computation
The Modbus communication system operates on 1 second intervals.
Within those intervals, data treatment like integration (whether analog or
digital) and servo control, are embedded in the module firmware. The SCB
receives the 1 second updates for higher level “software” processing and
control via SNMP registers, some of which is interfaced with the front
panel Graphical User Interface (GUI).
In addition to the operating registers, the 48iQ stores a historical data log
in a MySQL database. The memory is provided on the same micro SD card
where the operating software resides, for which there is capability to store
up to a year of data at 1 minute intervals. Chapter 3, “
Operation
” describes
how this database is accessed and used including external memory
downloads.
The 48iQ can be configured with an optional DMC based Oxygen (O
2
)
Sensor. This sensor is a paramagnetic sensor for O
2
concentration
measurement and CO correction. This option allows the user to correct the
CO readings for the amount of oxygen in the sample. Selectable O
2
concentrations can be used as the correction factor.
Oxygen Sensor
(optional)
Содержание 117080-00
Страница 1: ...48iQ Instruction Manual Carbon Monoxide Analyzer 117080 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 28: ......
Страница 36: ...Operation Instrument Display 3 8 48iQ Instruction Manual Thermo Scientific ...
Страница 154: ......
Страница 165: ...Calibration Calibration Thermo Scientific 48iQ Instruction Manual 4 11 ...
Страница 174: ...Maintenance Pump Rebuilding 5 4 48iQ Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 184: ......
Страница 192: ...Servicing Fan Replacement 7 8 48iQ Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 220: ...Servicing DMC Optical Bench 7 36 48iQ Instruction Manual Thermo Scientific Figure 7 40 Filter Wheel ...
Страница 230: ...Servicing DMC Optical Bench 7 46 48iQ Instruction Manual Thermo Scientific Figure 7 48 IR Source Cable Disconnect ...
Страница 252: ......
Страница 278: ......
Страница 290: ......
Страница 291: ......