Servicing
Optional Zero Air Scrubber Replacement
7-50
48iQ Instruction Manual
Thermo Scientific
Use the following procedure to replace the zero air scrubber.
Equipment required:
Hex drive, 9/16
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
Figure 2–1
).
2.
Disconnect plumbing.
3.
Unfasten four #8-32 socket cap head screws.
Figure 7–53.
Zero Air Scrubber Replacement
4.
Replace zero air scrubber and reinstall in reverse order.
Optional Zero Air
Scrubber