151
functions, not just for the pulse waveforms. If a different
function is chosen and the current period is out of range, the
period will be automatically adjusted to suit the new function.
Syntax
SOURce[1|2]:PULSe:PERiod
{<seconds>|MINimum|MAXimum}
Example
SOUR1:PULS:PER MIN
Sets the period to the minimum time allowed.
Query Syntax
SOURce[1|2]:PULSe:PERiod? [MINimum|MAXimum]
Return
Parameter
<seconds>
40ns~2000s
Example
SOUR1:PULS:PER?
+1.0000E+01
The period is set to 10 seconds.
8-8-2. SOURce[1|2]:PULSe:WIDTh
Set
Query
Description
Sets or queries the pulse width. The default pulse width is
100us.
The minimum pulse width is affected by the period time. If the
period is over 20 or 200 seconds, then the minimum pulse
width is 1us and 10us, respectively.
Pulse width is defined as the time from the rising to falling
edges (at a threshold of 50%).
Note
The pulse width cannot be less than the edge time times 1.6.
Pulse Width > 1.6 * Edge Time
The pulse width must be less than the period minus the edge
time (x1.6).
Pulse Width < Period
– (1.6 *Edge Time)
Syntax
SOURce[1|2]:PULSe:WIDTh
{<seconds>|MINimum|MAXimum}
Example
SOUR1:PULS:WIDT MAX
Sets the pulse width to the maximum allowed.
Query Syntax
SOURce[1|2]:PULSe:WIDTh? [MINimum|MAXimum]
Return
Parameter
<seconds>
20 ns ~ 1999.9 seconds
Содержание FGX-2220
Страница 1: ...B71 0406 01 INSTRUCTION MANUAL ARBITRARY FUNCTION GENERATOR FGX 2220...
Страница 44: ...31 3 1 5 Selecting a Noise Waveform Panel Operation 1 Press the Waveform key Waveform 2 Press F5 Noise Noise F5...
Страница 98: ...85...
Страница 137: ...124...
Страница 227: ...214 9 2 External Dimensions Figure...
Страница 229: ...7F Towa Fudosan Shin Yokohama Bldg 2 18 7 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...