181
Note
If the trigger source is set to immediate, the product of the
burst period and waveform frequency must be greater than
the burst count:
Burst Period X Waveform frequency > burst count
If the burst count is too large, the burst period will
automatically be increased and a “Settings conflict” error will
be generated.
Only sine and square waves are allowed infinite burst above
25 MHz.
Syntax
SOURce[1|2]:BURSt:NCYCles{< # cycles>
|INFinity|MINimum |MAXimum}
Parameter
<# cycles>
1~65535 cycles.
INFinity
Sets the number to continuous.
MINimum
Sets the number to minimum allowed.
MAXimum
Sets the number to maximum allowed.
Example
SOUR1:BURS:NCYCl INF
Sets the number of burst cycles to continuous (infinite).
Query Syntax
SOURce[1|2]:BURSt:NCYCles? [MINimum|MAXimum]
Return Parameter <NR3>
Returns the number of cycles.
INF
INF is returned if the number of cycles is
continuous.
Example
SOUR1:BURS:NCYC?
+1.0000E+02
The burst cycles are set to 100.
8-15-5. SOURce[1|2]:BURSt:INTernal:PERiod
Set
Query
Description
Sets or queries the burst period. Burst period settings are only
applicable when the trigger is set to immediate. The default
burst period is 10ms.
During manual triggering, external triggering or Gate burst
mode, the burst period settings are ignored.
Содержание FGX-2220
Страница 1: ...B71 0406 01 INSTRUCTION MANUAL ARBITRARY FUNCTION GENERATOR FGX 2220...
Страница 44: ...31 3 1 5 Selecting a Noise Waveform Panel Operation 1 Press the Waveform key Waveform 2 Press F5 Noise Noise F5...
Страница 98: ...85...
Страница 137: ...124...
Страница 227: ...214 9 2 External Dimensions Figure...
Страница 229: ...7F Towa Fudosan Shin Yokohama Bldg 2 18 7 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...