![Tektronix 2440 Скачать руководство пользователя страница 297](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860297.webp)
Maintenance— 2440 Service
Table 6-6 (cont)
The check is set up by first doing a COLD START (to again set the calibration constants to known
values), and then applying a 4-division signal to the CH 1 and CH 2 vertical inputs (or to the bad
channel if only one is bad). Observe the signal on the crt, if possible; otherwise, use a test scope
to probe the vertical signal path to check for correct response. Since the Preamp outputs are not
accessible, use the output of the Peak Detectors to verify the signal through the Preamp. The side
3 Peak Detector output for CH 1 may be checked on R441 and R540; and for the CH 2 side 3
output, check at R244 and R 341.
Use either the Front Panel Controls or the Force DAC test to check the VARIABLE GAIN and
VERTICAL POSITION. Balance may be varied only using the Force DAC function. Varying the
channel balance should appear as a dc offset change to the vertical signal level. Prior to making
any adjustments after a COLD START (either with the Front Panel controls or the Force DAC
function), the signal at the output of the Peak Detectors should have a + 8 .7 V dc level with an ac
signal (replica of the input signal) of approximately 0.5 V peak-to-peak.
5.
The Preamplifier operating mode is set by a serial data word sent from the System
m
P. The CD
input, pin 22, is the serial data input. A TTL-level logic swing should be present on this line when
ever the Attenuators, Preamps, or A/B Trigger Generator operating modes are being set up. If
there are no FAIL flags under these major test categories, the CD circuitry is most likely function
ing properly. The CC input, pin 23, is the control clock input, and it should have a TTL-level logic
swing on it only when the particular hybrid, in this case the Preamp hybrid, is being set up. Check
that this line is high initially and pulses LO eight times for each Preamp load cycle. (The eight
pulses may be separated into several groups of pulses.)
6.
The Preamps have bypassed and decoupled voltage supplies. Check that the bypassing com
ponents in series with the power supplies are not open. Also check that the dc bias voltages at
the Preamp are approximately the levels indicated on the schematic diagram in the service
manual.
7.
If the Preamp hybrid itself is suspected of being defective and the other channel is fully functional,
swap Preamp hybrids between channels to see if the problem moves to the other channel. If so,
replace the defective Preamp. If not, check the hybrid mount connections of the defective channel
for corrosion or contamination that may be causing a poor contact.
6 -8 9
Содержание 2440
Страница 4: ......
Страница 12: ......
Страница 37: ...Specification 2440 Service Dimensions are in inches mm 4918 39 1 25 Figure 1 1 Dimensional drawing ...
Страница 38: ......
Страница 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Страница 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Страница 123: ...Theory of Operation 2440 Service SOURCE ACCEPTOR 4 9 1 7 1 2 3 81 Figure 3 11 GPIB data flow diagram ...
Страница 130: ...Theory of Operation 2440 Service 3 88 ...
Страница 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Страница 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Страница 208: ......
Страница 322: ......
Страница 419: ......
Страница 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Страница 427: ......
Страница 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Страница 429: ......
Страница 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Страница 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Страница 439: ......
Страница 445: ......
Страница 446: ...2440 Service WAVEFORMS FOR DIAGRAM 7 I I f f f NtITOT CHI r K I It MVIHi TRIGGER ON FALLING EDGE OF U832 PIN 8 MORE ...
Страница 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Страница 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Страница 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Страница 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Страница 468: ...2440 Service W A VE FO R M S FOR DIAG RAM 17 TEST SCOPE TRIGGERED ON 126 6330 46 WAVEFORMS FOR DIAGRAM 17 ...
Страница 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Страница 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Страница 490: ...YES ...
Страница 492: ......
Страница 494: ......
Страница 499: ...YES ...
Страница 501: ......
Страница 502: ...2440 Service FIG 1 C A B IN E T ...
Страница 503: ......
Страница 510: ...2440 Service SISSVHO Z Old ...
Страница 511: ......
Страница 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Страница 513: ......
Страница 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Страница 518: ...2440 Service ...
Страница 519: ...FIG 4 L V H V PWR SUPPLY ...
Страница 520: ...FIG 5 ACCESSORIES 2440 Service ...
Страница 521: ......
Страница 523: ......
Страница 529: ......
Страница 530: ......
Страница 531: ......