![Tektronix 2440 Скачать руководство пользователя страница 175](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860175.webp)
y.
Push PRGM. V ER IFY —That the BELL rings and the
sequence loops back to display step 1 of the sequence.
z.
Connect the STEP COM PLETE output BNC (located on
rear panel) to the banana plug Inputs of a Digital
Voltmeter (DMM) via a 50
Cl
Coaxial Cable and a BNC
Female-to-Dual Banana Adapter. When connecting the
BNC Female-to-Dual Banana Adapter to the DMM, put
the side with the bump marked “G N D ” to the LOW or (-)
input jack.
aa. Set the DMM to the 20 D C VOLTS range. C H E C K -T h a t
the DMM reading is < 0.5 V.
ab. Push PRGM to advance to sequence step 2. C H E C K -
That the DMM reading momentarily jumps to a level
^ 2.5 V and < 3.5 V before returning to the level
measured in subpart aa.
ac. Move the 50 n
Coaxial Cable from the STEP
COMPLETE output to the S EQ U E N C E O U T output
BNC. C H E C K -T h a t the DMM reading is ^ 2.5 V and
< 3.5 V.
ad. Push PRGM once to advance to sequence step 3. Wait
until step 3 is loaded and then push PRGM again to
advance to step 4.
ae. C H E C K —That the DMM reading is < 0.5 V.
af. Disconnect the
50 f i
Coaxial Cable from the
S EQ U E N C E O U T output BNC.
ag. Connect the square wave output of a generator (such as
Item 19) capable of outputting nominal TTL levels to the
S EQ U E N C E IN input BNC via a 50 f l Coaxial Cable.
Set the output frequency of the generator to 10 Hz.
ah. V E R IF Y —That the scope continuously loops through
sequencer steps
1
to 4
in response to the
generator input.
ai. Disconnect the
50 f l
Coaxial Cable from the
S EQ U E N C E IN BNC.
aj. Push
EXIT.
V E R IF Y —That
the
RECALL
menu
is returned.
ak. Push EXIT. V E R IF Y —That the main AUTOSTEP
S EQ UEN CER menu is returned.
7. GPIB Functionality Verification.
a.
Press the instrument’s POWER button twice to power
the instrument O FF and then ON .
b.
V ER IFY —That all three GPIB STATUS lights illuminate
during the instrument's power-up sequence.
Performance Check and Functional Verification Procedure
2440 Service
c.
V ER IFY —The GPIB STATUS SRQ light is still illumi
nated when the power-up sequence is finished.
d.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL S E T U P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
e.
Push SETUP O U TPU T (the button to the lower right of
S EC /D IV control) and press the menu button labeled
. SETUP (menu will change).
f.
Press the O U TPU T SETUP menu button labeled M ODE
to display that menu.
g.
Set T /L on. V E R IF Y —That the ADDR light is off.
h.
Set L/ONLY on. V E R IF Y —That the ADDR light is on.
i.
Set T/ONLY on. V E R IF Y —That the ADDR light remains
on.
j.
Push SETUP O U TPU T and press the menu button
labeled SETUP.
k.
Press the OU TPU T SETU P menu button labeled M O DE
to display that menu.
l
.
Set T /L on.
m. Push SETUP O U TPU T and press the menu button
labeled SETUP.
n.
Press the menu button labeled ADDR to select that
menu.
o.
Press the menu button labeled f or J. to set the GPIB
A DD RESS to 1. The t increments the address and the
j decrements it.
p.
Press SETUP O U T P U T and press the menu button
labeled SETUP. Press the menu button labeled TERM
(menu will change).
NOTE
At this point In this step, the use of a GPIB
compatible Controller is necessary to send
commands or queries over the GPIB to the 2440
and receive responses back from the 2440.
Due to the variety of controller types, the
following parts of this step only give some
examples of commands you can have your
controller send to the 2440 over the GPIB Bus to
verify the GPIB Interface in this Instrument.
q.
Set either EOI or LF/EOI on accoring to the specifica
tion of the GPIB Controller.
r.
Turn on the GPIB Controller and enter a program that
can deliver commands and queries to, as well as
receive response from, the scope. Below are some
limited commands your program could perform using
the GPIB controller:
4-33
Содержание 2440
Страница 4: ......
Страница 12: ......
Страница 37: ...Specification 2440 Service Dimensions are in inches mm 4918 39 1 25 Figure 1 1 Dimensional drawing ...
Страница 38: ......
Страница 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Страница 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Страница 123: ...Theory of Operation 2440 Service SOURCE ACCEPTOR 4 9 1 7 1 2 3 81 Figure 3 11 GPIB data flow diagram ...
Страница 130: ...Theory of Operation 2440 Service 3 88 ...
Страница 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Страница 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Страница 208: ......
Страница 322: ......
Страница 419: ......
Страница 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Страница 427: ......
Страница 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Страница 429: ......
Страница 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Страница 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Страница 439: ......
Страница 445: ......
Страница 446: ...2440 Service WAVEFORMS FOR DIAGRAM 7 I I f f f NtITOT CHI r K I It MVIHi TRIGGER ON FALLING EDGE OF U832 PIN 8 MORE ...
Страница 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Страница 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Страница 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Страница 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Страница 468: ...2440 Service W A VE FO R M S FOR DIAG RAM 17 TEST SCOPE TRIGGERED ON 126 6330 46 WAVEFORMS FOR DIAGRAM 17 ...
Страница 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Страница 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Страница 490: ...YES ...
Страница 492: ......
Страница 494: ......
Страница 499: ...YES ...
Страница 501: ......
Страница 502: ...2440 Service FIG 1 C A B IN E T ...
Страница 503: ......
Страница 510: ...2440 Service SISSVHO Z Old ...
Страница 511: ......
Страница 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Страница 513: ......
Страница 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Страница 518: ...2440 Service ...
Страница 519: ...FIG 4 L V H V PWR SUPPLY ...
Страница 520: ...FIG 5 ACCESSORIES 2440 Service ...
Страница 521: ......
Страница 523: ......
Страница 529: ......
Страница 530: ......
Страница 531: ......