![Tektronix 2440 Скачать руководство пользователя страница 176](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860176.webp)
Performance C heck and Functional Verification Procedure
2440 Service
1.
Ask for the user to enter the 2440’s GPIB address.
(This was set to 1 in part o.)
2.
Have the user send the command RQS ON.
3.
Have the user send the command LOCK ON to
verify the LOCK light illuminates. Next send the
command LOCK OFF to the 2440 and verify that the
LOCK light is extinguished,
4.
Have the user send commands with the GPIB
Controller to set up the front panel controls. Some
examples are:
VM Ode ADD:ON
C H I VOLts:1,COUpling:GND, INVert:ON
BWLimit TWEnty
5.
Have the user send a query to the 2440 (i.e.
VMOde?) to verify the 2440 sends back the
response.
NOTE
Refer to the 2440 Programmers Reference
Guide for sample programs for several different
Controllers.
s.
Connect the GPIB Controller to the oscilloscope’s
rear-panel GPIB C O N N E C TO R using the GPIB cable.
t.
Run the program entered for subpart r.
u.
After you have sent commands over the GPIB to the
2440, press STATUS to verify that the front panel
STATUS readout indicates the control setting changes
sent over the GPIB Controller.
v.
Press the M ENU O FF/E X TEN D E D FU N C TIO N S button.
w.
Disconnect the test setup.
8. Check A TRIGGER and RECORD TRIGGER
Outputs for Logic Polarity and Minimum
HI/LO (50 n loads).
a.
Recall the Initial Front-Panel Setup, labeled “FPNL"
(see step i in “INITIAL S E TU P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
Select: CH 1 C O U PLIN G /IN VER T
Set:
50 f l ON | OFF
OFF
Select: CH 2 C O U PLIN G /IN VER T
Set:
50 O ON | OFF
OFF
Set:
CH 1 VOLTS/DIV
200 mV
CH 2 VOLTS/DIV
200 mV
Select: TRIGGER S O U RC E
Set:
LINE
On
b.
Connect the RECORD TRIGG ER O U TPU T (rear panel)
to the CH 1 OR X input connector via a 50
Cl
Coaxial
Cable and a 50 O Termination.
c.
Connect the A TRIGGER O U TPU T (rear panel) to the
CH 2 OR Y input connector via a 50
£1
Coaxial Cable
and a 50 f l Termination.
d.
Using the CH 1 and CH 2 VERTICAL POSITION
controls, position the CH 1 waveform to the top-half of
the screen and the CH 2 to the bottom-half for easy
viewing.
e.
C H E C K -T h a t both of the waveforms are displayed
with their falling edges aligned to the Trigger Point
Indicator (a small “T ” riding on each waveform).
f.
Select CURSORS FUN CTIO N and set VOLTS on.
g.
Select CURSOR UNITS and set A | ABS to ABS.
h.
Use the CURSOR/DELAY control to align the Voltage
cursor to the top flat portion of the CH 1 waveform.
i.
C H E C K -T h a t the Cursor Readout indicates a voltage
^ 450 mV.
j.
Align the Voltage cursor to the bottom flat portion of the
C H 1 waveform.
k.
C H E C K -T h a t the Cursor Readout indicates a voltage
< 150 mV.
l
.
Press the CURSO R FU N C TIO N button twice to display
the Attach Cursors menu and set CH 2 on for the
displayed menu.
m. Repeat parts h through k, aligning the cursor to the
CH 2 waveform instead of the CH 1.
n.
Disconnect the test setups.
4-34
Содержание 2440
Страница 4: ......
Страница 12: ......
Страница 37: ...Specification 2440 Service Dimensions are in inches mm 4918 39 1 25 Figure 1 1 Dimensional drawing ...
Страница 38: ......
Страница 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Страница 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Страница 123: ...Theory of Operation 2440 Service SOURCE ACCEPTOR 4 9 1 7 1 2 3 81 Figure 3 11 GPIB data flow diagram ...
Страница 130: ...Theory of Operation 2440 Service 3 88 ...
Страница 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Страница 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Страница 208: ......
Страница 322: ......
Страница 419: ......
Страница 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Страница 427: ......
Страница 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Страница 429: ......
Страница 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Страница 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Страница 439: ......
Страница 445: ......
Страница 446: ...2440 Service WAVEFORMS FOR DIAGRAM 7 I I f f f NtITOT CHI r K I It MVIHi TRIGGER ON FALLING EDGE OF U832 PIN 8 MORE ...
Страница 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Страница 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Страница 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Страница 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Страница 468: ...2440 Service W A VE FO R M S FOR DIAG RAM 17 TEST SCOPE TRIGGERED ON 126 6330 46 WAVEFORMS FOR DIAGRAM 17 ...
Страница 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Страница 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Страница 490: ...YES ...
Страница 492: ......
Страница 494: ......
Страница 499: ...YES ...
Страница 501: ......
Страница 502: ...2440 Service FIG 1 C A B IN E T ...
Страница 503: ......
Страница 510: ...2440 Service SISSVHO Z Old ...
Страница 511: ......
Страница 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Страница 513: ......
Страница 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Страница 518: ...2440 Service ...
Страница 519: ...FIG 4 L V H V PWR SUPPLY ...
Страница 520: ...FIG 5 ACCESSORIES 2440 Service ...
Страница 521: ......
Страница 523: ......
Страница 529: ......
Страница 530: ......
Страница 531: ......